PIXEL-BASED CORRECTION METHOD FOR GAFCHROMIC (R) EBT FILM DOSIMETRYopen access

Authors
Jeong, Hae SunHan, YoungyihKum, OyeonKim, Chan HyeongJu, Sang GyuShin, Jung SukKim, Jin SungPark, Joo Hwan
Issue Date
Dec-2010
Publisher
KOREAN NUCLEAR SOC
Keywords
EBT Film; Flat-Bed Scanner; Non-Uniform Dose Correction
Citation
NUCLEAR ENGINEERING AND TECHNOLOGY, v.42, no.6, pp.670 - 679
Indexed
SCIE
SCOPUS
KCI
Journal Title
NUCLEAR ENGINEERING AND TECHNOLOGY
Volume
42
Number
6
Start Page
670
End Page
679
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/173361
DOI
10.5516/NET.2010.42.6.670
ISSN
1738-5733
Abstract
In this paper, a new approach using a pixel-based correction method was developed to fix the non-uniform responses of flat-bed type scanners used for radiochromic film dosimetry. In order to validate the method's performance, two cases were tested: the first consisted of simple dose distributions delivered by a single port; the second was a complicated dose distribution composed of multiple beams. In the case of the simple individual dose condition, ten different doses, from 8.3 cGy to 307.1 cGy, were measured, horizontal profiles were analyzed using the pixel-based correcton method and compared with results measured by an ionization chamber and results corrected using the existing correction method. A complicated inverse pyramid dose distribution was made by piling up four different field shapes, which were measured with GAFCHROMIC (R) EBT film and compared with the Monte Carlo calculation; as well as the dose distribution corrected using a conventional method. The results showed that a pixel-based correction method reduced dose difference from the reference measurement down to 1% in the flat close distribution region or 2 mm in a steep dose gradient region compared to the reference data, which were ionization chamber measurement data for simple cases and the MC computed data for the complicated case, with an exception for very low doses of less than about 10 cGy in the simple case. Therefore, the pixel-based scanner correction method is expected to enhance the accuracy of GAFCHROMIC (R) EBT film dosimetry, which is a widely used tool for two-dimensional dosimetry.
Files in This Item
Go to Link
Appears in
Collections
서울 공과대학 > 서울 원자력공학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Chan Hyeong photo

Kim, Chan Hyeong
COLLEGE OF ENGINEERING (DEPARTMENT OF NUCLEAR ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE