Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

위전도 측정을 위한 전극간 부착거리에 관한 연구Influences of Inter-electrode Distance on Electrogastrography Measurements

Other Titles
Influences of Inter-electrode Distance on Electrogastrography Measurements
Authors
한완택송인호김인영
Issue Date
Aug-2009
Publisher
대한의용생체공학회
Keywords
Electrogastrography; Surface electrodes; bipolar recording; surface potential; fat thickness; Electrogastrography; Surface electrodes; bipolar recording; surface potential; fat thickness
Citation
의공학회지, v.30, no.4, pp.341 - 346
Indexed
KCI
Journal Title
의공학회지
Volume
30
Number
4
Start Page
341
End Page
346
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/176311
ISSN
1229-0807
Abstract
Cutaneous electrogastrography is the measurement of electrical activity of the stomach on the abdominal surface. The validity of cutaneous electrogastrography is dependent upon the quality of the recording technique. The locations of electrodes are an important issue. We examined the influences of the inter-electrode distance of bipolar leads on electrogastrography measurements. The sensitivity distributions of EGG leads were calculated based on a 2D body fat model and evaluated according to the region of interest sensitivity ratio (ROISR). We simulated the ROISR of the inter-electrode distance in relation to various body fat thicknesses. The distance between the electrodes was proportional to the distance between the ROI and the surface of the abdomen. The results imply that inter-electrode distance can be applied in electrogastrography according to human body fat thickness.
Files in This Item
Go to Link
Appears in
Collections
서울 의과대학 > 서울 의공학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, In Young photo

Kim, In Young
COLLEGE OF MEDICINE (DEPARTMENT OF BIOMEDICAL ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE