Development of a beta gauge system for a fabric density measurement
- Authors
- Kim, Han Soo; Park, Se Hwan; Ha, Jang Ho; Song, Tae Young; Cho, Seung Yeon; Kim, Yong Kyun
- Issue Date
- Jul-2009
- Publisher
- Pergamon Press Ltd.
- Keywords
- Beta gauge; Kr-85; Ionization chamber; Density measurement; NDT
- Citation
- Applied Radiation and Isotopes, v.67, no.7-8, pp 1213 - 1215
- Pages
- 3
- Indexed
- SCIE
SCOPUS
- Journal Title
- Applied Radiation and Isotopes
- Volume
- 67
- Number
- 7-8
- Start Page
- 1213
- End Page
- 1215
- URI
- https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/176573
- DOI
- 10.1016/j.apradiso.2009.02.015
- ISSN
- 0969-8043
1872-9800
- Abstract
- A beta gauge system was designed and fabricated for a fabric density measurement with high sensitivity in real time. An ionization chamber and 3.7 GBq Kr-85 was chosen as a beta radiation sensor and a beta source. An ionization chamber, which had a 0.5 mu m-thick aluminized mylar film as an incident window, was constructed with a guard electrode structure in consideration of minimizing leakage currents. Xe gas was chosen as a filling gas to increase sensitivity of an ionization chamber. Linearity of the ionization currents against the densities was 0.9894 in the selected measurement range. And the deviation of the ionization currents at a standard density was below 0.02%. The ranges of the density measurement and the active area were from 20 to 999 g/cm(2) with +/- 1% accuracy and 40 mm diameters. The fabricated beta gauge system can be applied to fields such as in the paper, film, and plastic production industries.
- Files in This Item
-
Go to Link
- Appears in
Collections - 서울 공과대학 > 서울 원자력공학과 > 1. Journal Articles

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.