Cited 0 time in
Simulation Analysis of Ion Chambers for Area Radiation Monitoring System
| DC Field | Value | Language |
|---|---|---|
| dc.contributor.author | Song, Tae-Yung | - |
| dc.contributor.author | Kim, Han Soo | - |
| dc.contributor.author | Park, Se-Hwan | - |
| dc.contributor.author | Ha, Jang Ho | - |
| dc.contributor.author | Kim, Yong-Kyun | - |
| dc.date.accessioned | 2022-12-21T02:49:18Z | - |
| dc.date.available | 2022-12-21T02:49:18Z | - |
| dc.date.issued | 2008-06 | - |
| dc.identifier.issn | 0022-3131 | - |
| dc.identifier.issn | 1881-1248 | - |
| dc.identifier.uri | https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/178545 | - |
| dc.description.abstract | KAERI (Korea Atomic Energy Research Institute) has been developing ion chambers which are used for an area radiation monitoring system. Those ion chambers will be placed in the area where it is necessary to monitor radiation dose all the time. The areas are nuclear plant site, research reactor site, radiation waste disposal site etc.. KAERI is developing a few different types of ion chambers for the above-mentioned purposes. One is an ion chamber having cylindrical shape with the active volume of similar to 1 l. Another one is also a cylindrical type ion chamber with the active volume of similar to 10 l. The smaller ion chamber of those 2 detectors will be installed in the place where a compact detector is required. Those ion chambers can be filled with gases such as air or Ar or Xe. The characteristics of those two ion chambers can be analyzed by simulation study. The simplified design of air-filled detectors (1.2 l and 13.7 l) was made and simulation study was performed by using the code EGSnrc. The energy of parallel radiation beam was varied from 50 keV to 1.5 MeV to analyze the expected ionization currents and dose as function of beam energy. The 100 keV and I MeV point source beams were adopted to investigate the angular dependence of radiation dose absorbed in the detector. We also investigated the change of characteristics of the 1.2 l ion chamber when the type of filled gas is changed from air. The radiation dose was obtained for the cases of Ar and Xe. The same study was also performed when the pressurized air (2 bar) was filled in the detector. | - |
| dc.format.extent | 4 | - |
| dc.language | 영어 | - |
| dc.language.iso | ENG | - |
| dc.publisher | Atomic Energy Society of Japan/Nihon Genshiroku Gakkai | - |
| dc.title | Simulation Analysis of Ion Chambers for Area Radiation Monitoring System | - |
| dc.type | Article | - |
| dc.publisher.location | 영국 | - |
| dc.identifier.doi | 10.1080/00223131.2008.10875899 | - |
| dc.identifier.scopusid | 2-s2.0-84912008553 | - |
| dc.identifier.wosid | 000267196000129 | - |
| dc.identifier.bibliographicCitation | Journal of Nuclear Science and Technology, pp 496 - 499 | - |
| dc.citation.title | Journal of Nuclear Science and Technology | - |
| dc.citation.startPage | 496 | - |
| dc.citation.endPage | 499 | - |
| dc.type.docType | Article; Proceedings Paper | - |
| dc.description.isOpenAccess | N | - |
| dc.description.journalRegisteredClass | scie | - |
| dc.description.journalRegisteredClass | scopus | - |
| dc.relation.journalResearchArea | Nuclear Science & Technology | - |
| dc.relation.journalWebOfScienceCategory | Nuclear Science & Technology | - |
| dc.subject.keywordPlus | Computer simulation | - |
| dc.subject.keywordPlus | Intelligent systems | - |
| dc.subject.keywordPlus | Ionization of gases | - |
| dc.subject.keywordPlus | Ions | - |
| dc.subject.keywordPlus | Monitoring | - |
| dc.subject.keywordPlus | Monte Carlo methods | - |
| dc.subject.keywordPlus | Nuclear reactors | - |
| dc.subject.keywordPlus | Radiation | - |
| dc.subject.keywordPlus | Radioactive waste disposal | - |
| dc.subject.keywordPlus | Waste disposal | - |
| dc.subject.keywordAuthor | area radiation monitoring | - |
| dc.subject.keywordAuthor | Monte Carlo simulation | - |
| dc.subject.keywordAuthor | EGSnrc | - |
| dc.subject.keywordAuthor | ion chamber | - |
| dc.subject.keywordAuthor | gas radiation detector | - |
| dc.identifier.url | This work was performed under the nuclear research and development program by Ministry of Science and Technology of Korea, and also supported by the Innovative Technology Center for Radiation Safety (iTRS). | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
222, Wangsimni-ro, Seongdong-gu, Seoul, 04763, Korea+82-2-2220-1366
COPYRIGHT © 2024 HANYANG UNIVERSITY.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.