A position control method for a robotically assisted magnetic navigation system to improve the pushability of a magnetic catheter by maximizing magnetic forceopen access

Authors
Lee, D.Jung, E.Kwon, J.Jang, G.
Issue Date
Mar-2023
Publisher
AIP Publishing
Citation
AIP ADVANCES, v.13, no.3, pp.1 - 4
Indexed
SCIE
SCOPUS
Journal Title
AIP ADVANCES
Volume
13
Number
3
Start Page
1
End Page
4
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/184928
DOI
10.1063/9.0000390
ISSN
2158-3226
Abstract
We propose a novel position control method (PCM) for the Robotically Assisted MAgnetic Navigation (RAMAN) system to improve the pushability of the magnetic catheter (MC) by maximizing the magnetic force. We expressed the magnetic force acting on the magnetic catheter as the position and the current flowing through each of the electromagnets of the RAMAN system. Next, we formulated a PCM as an optimization problem to maximize the MF. From the proposed PCM, we can determine the position that generates the maximum MF in the desired direction within the region of interest of the RAMAN as well as the required applied current of each electromagnet. Finally, we performed a navigation experiment for the MC along the aorta to the right coronary artery in a cardiac vascular phantom model, and we showed that the proposed PCM can navigate through the vascular phantom effectively without buckling of the MC.
Files in This Item
Appears in
Collections
서울 공과대학 > 서울 기계공학부 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Jang, Gun Hee photo

Jang, Gun Hee
COLLEGE OF ENGINEERING (SCHOOL OF MECHANICAL ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE