초음파를 이용한 팬텀 근육 내 주입한 초음파 젤 부피 계측: 예비연구Ultrasonographic Measurements of the Intramuscular Fluid Collection Using a Phantom: A Preliminary Study
- Other Titles
- Ultrasonographic Measurements of the Intramuscular Fluid Collection Using a Phantom: A Preliminary Study
- Authors
- 이희제; 안새미; 류정아; 박태수
- Issue Date
- Aug-2016
- Publisher
- 대한정형외과학회
- Keywords
- 근육내 젤 저류; 계측; 관찰자 간 일치도; 초음파; intermuscular gel collection; measurement; interobserver reliability; ultrasonography
- Citation
- 대한정형외과학회지, v.51, no.4, pp 315 - 319
- Pages
- 5
- Indexed
- KCI
- Journal Title
- 대한정형외과학회지
- Volume
- 51
- Number
- 4
- Start Page
- 315
- End Page
- 319
- URI
- https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/194128
- DOI
- 10.4055/jkoa.2016.51.4.315
- ISSN
- 1226-2102
2005-8918
- Abstract
- Purpose
The purpose of this study is to check the intermuscular fluid collection in a phantom using ultrasound after comparison with the original injected volume of gel and to evaluate interobserver reliability.
Materials and Methods
One orthopedic surgeon injected ultrasonographic gel into the muscle of slaughtered cow meat using a syringe and needle, and recorded the injected volume. The injected volume of gel ranged from 1 ml to 5 ml, and total number of trials was 30 times. The volume was calculated using ultrasound by two experienced radiologists. The ultrasonographic measured volme was assessed and interobserver variability was measured.
Results
The mean ultrasonographic measured volume was 71.35% of the original injected volume of gel. Thus, the measured volume using ultrasound showed a tendency to underestimate the injected volume into the muscle. For evaluation of interobserver reliability, interclass correlation using SPSS and consistency using Bland and Altman plot were calculated. Interclass correlation was 0.95, and Bland and Altman plot showed that the differences of mean volume measured by two observers were within ±1.96 standard deviation over 95%. Therefore, it showed good interoberver reliability.
Conclusion
Ultrasonographic measurement of the intramuscular fluid collection in a phantom showed mean 71.35% of the original injected volume of gel, and it tended to be underestimated in proportion to the increased volume of intermuscular gel. This study showed good interobserver reliability.