Detailed Information

Cited 15 time in webofscience Cited 8 time in scopus
Metadata Downloads

Fabrication of Vacuum-Sealed Capacitive Micromachined Ultrasonic Transducer Arrays Using Glass Reflow Processopen access

Authors
Nguyen Van ToanHahng, ShimSong, YunheubOno, Takahito
Issue Date
May-2016
Publisher
MDPI
Keywords
capacitive micromachined ultrasonic transducer; glass reflow process; anodic bonding; medical imaging; non-destructive measurement; chemical sensing
Citation
MICROMACHINES, v.7, no.5, pp.1 - 9
Indexed
SCIE
SCOPUS
Journal Title
MICROMACHINES
Volume
7
Number
5
Start Page
1
End Page
9
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/23116
DOI
10.3390/mi7050076
ISSN
2072-666X
Abstract
This paper presents a process for the fabrication of vacuum-sealed capacitive micromachined ultrasonic transducer (CMUT) arrays using glass reflow and anodic bonding techniques. Silicon through-wafer interconnects have been investigated by the glass reflow process. Then, the patterned silicon-glass reflow wafer is anodically bonded to an SOI (silicon-on-insulator) wafer for the fabrication of CMUT devices. The CMUT 5 x 5 array has been successfully fabricated. The resonant frequency of the CMUT array with a one-cell radius of 100 mu m and sensing gap of 3.2 mu m (distance between top and bottom electrodes) is observed at 2.84 MHz. The Q factor is approximately 1300 at pressure of 0.01 Pa.
Files in This Item
Appears in
Collections
서울 공과대학 > 서울 융합전자공학부 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Song, Yun Heub photo

Song, Yun Heub
COLLEGE OF ENGINEERING (SCHOOL OF ELECTRONIC ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE