Fixed volume sequential standard addition calibration: Value assignment of impurities in zero gas
- Authors
- Brown, Richard J. C.; Brewer, Paul J.; Kim, Ki-Hyun
- Issue Date
- May-2017
- Publisher
- Elsevier BV
- Keywords
- standard addition; calibration; bias; zero gas; matrix
- Citation
- Chemometrics and Intelligent Laboratory Systems, v.164, pp 52 - 55
- Pages
- 4
- Indexed
- SCI
SCIE
SCOPUS
- Journal Title
- Chemometrics and Intelligent Laboratory Systems
- Volume
- 164
- Start Page
- 52
- End Page
- 55
- URI
- https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/4224
- DOI
- 10.1016/j.chemolab.2017.03.012
- ISSN
- 0169-7439
1873-3239
- Abstract
- A fixed volume case of sequential standard addition calibration (S-SAC) for the value assignment of impurities in zero gas is described. A mathematical description of this technique has been derived and has been shown to exhibit a similar systematic bias during the extrapolation process to that seen for other S-SAC cases. The use of S-SAC in the gas phase has demonstrated that variation in the sample volume for S-SAC in the liquid phase is not the generalised cause of the bias experienced during extrapolation. Instead it is the variation in the volume of the original sample with respect to the overall volume of the mixture following the addition of standard. In addition, the requirement to perform a bias correction on the extrapolated values has been discussed and best practice solutions for correction proposed.
- Files in This Item
-
Go to Link
- Appears in
Collections - 서울 공과대학 > 서울 건설환경공학과 > 1. Journal Articles

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.