Detailed Information

Cited 0 time in webofscience Cited 1 time in scopus
Metadata Downloads

Impaction of a continuous glucose monitoring sensoropen access

Authors
Park, Kyong ChanChoi, Hwan Jun
Issue Date
Jul-2021
Publisher
대한성형외과학회
Keywords
Biosensor; Home blood glucose monitoring; Foreign bodies
Citation
Archives of Plastic Surgery, v.48, no.4, pp 392 - 394
Pages
3
Journal Title
Archives of Plastic Surgery
Volume
48
Number
4
Start Page
392
End Page
394
URI
https://scholarworks.bwise.kr/sch/handle/2021.sw.sch/18704
DOI
10.5999/aps.2021.00178
ISSN
2234-6163
2234-6171
Abstract
A 33-year-old man presented to the plastic surgery department for foreign body removal 1 month after the insertion of a continuous glucose monitoring (CGM) sensor (Dexcom G5) in the left upper arm. The patient had used the CGM system for 5 years, and the insertion was done in the usual manner. The entire sensor wire was visible on simple radiography and ultrasonography. In the operating room, and the sensor wire was identified in the intermuscular septum and removed. No foreign body reaction or inflammatory signs were found around the CGM, and the extracted wire measured 2.5 cm. Thus, it was assumed that the whole sensor wire was detached from the transmitter, not fractured. No remnant foreign body was observed on follow-up simple radiography.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Plastic Surgery > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Choi, Hwan jun photo

Choi, Hwan jun
College of Medicine (Department of Plastic Surgery)
Read more

Altmetrics

Total Views & Downloads

BROWSE