A Comparison of Central Corneal Thickness Measurements and Measurement Repeatability Using Three Imaging Modalities세 가지 기기를 이용하여 측정한 중심각막두께의 비교 및 반복성 비교

Other Titles
세 가지 기기를 이용하여 측정한 중심각막두께의 비교 및 반복성 비교
Authors
Woo, Sang EarnLee, Si Hyung
Issue Date
Feb-2021
Publisher
대한안과학회
Keywords
Anterior segment optical coherence tomography; Central corneal thickness; Dual rotating Scheimpflug camera; Ultrasound pachymetry
Citation
대한안과학회지, v.62, no.2, pp 184 - 192
Pages
9
Journal Title
대한안과학회지
Volume
62
Number
2
Start Page
184
End Page
192
URI
https://scholarworks.bwise.kr/sch/handle/2021.sw.sch/18996
DOI
10.3341/jkos.2021.62.2.184
ISSN
0378-6471
2092-9374
Abstract
Purpose: To compare and analyze central corneal thickness (CCT) measurements and measurement repeatability according to the corneal thickness obtained with anterior segment optical coherence tomography (AS-OCT), ultrasound pachymetry (USP), and dual Scheimpflug imaging (Galilei G4 (R)). Methods: CCT measurements obtained using AS-OCT, USP, and a dual rotating Scheimpflug camera in 120 eyes of 60 healthy subjects were compared and the repeatability was analyzed. Results: The average CCT measurements using AS-OCT, USP, and a dual rotating Scheimpflug camera were 511.23 +/- 37.16, 525.78 +/- 39.22, and 531.43 +/- 41.57 mu m, respectively. The CCT measurement using dual rotating Scheimpflug imaging was statistically significantly thicker than that measured using AS-OCT (p < 0.05). A significant linear correlation was observed among OCT, USP, and dual rotating Scheimpflug measurements (r > 0.900, p < 0.001). All three methods showed high repeatability (intraclass correlation coefficient >= 0.996) for intra-examiner CCT measurements. Conclusions: The CCT measurement results using the three methods were significantly correlated and showed high repeatability. However, the AS-OCT measurement was significantly thinner than that obtained using a dual rotating Scheimpflug camera. CCT measurement with a dual rotating Scheimpflug camera is a better alternative for USP than AS-OCT. Therefore, CCT should be interpreted in the context of the instrument used.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Ophthalmology > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Si Hyung photo

Lee, Si Hyung
College of Medicine (Department of Ophthalmology)
Read more

Altmetrics

Total Views & Downloads

BROWSE