Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Spray type과 Soot type 입자발생기별 나노입자 개수농도분포 측정특성 비교

Full metadata record
DC Field Value Language
dc.contributor.author김민식-
dc.contributor.author권지원-
dc.contributor.author정명철-
dc.contributor.author이진욱-
dc.date.available2018-05-10T07:49:18Z-
dc.date.created2018-04-17-
dc.date.issued2012-12-
dc.identifier.issn1226-2277-
dc.identifier.urihttp://scholarworks.bwise.kr/ssu/handle/2018.sw.ssu/13279-
dc.description.abstractParticulate matters(PM) that is generated by most diesel engine is regulated by the mass concentration measured by the conventional method it had been. Recently, Europe PMP (Particle Measurement Program) decided to start the regulation of vehicle's nano-sized particle number (PN) from the year of 2011 because of nano-particle's higher degree of harm to the human body. So firstly, the standard level of PN emission is introduced in the Euro 5/6 emissions regulation with a limit of 6x1011 per km for light duty vehicle. Also KPMP(Korea Particle Measurement Program) was organized to copy quickly international technical trend. In this paper, it was investigated the nano-sized PN measurement characteristics for different particle generators between spray type and soot type. And the difference ratio between particle generators, the characteristic of PN concentration, counting efficiency and linearity was analyzed. Then, we make conclusions as followed. When particle diameter is increased, counting efficiency of two generators is decreased. Also Secondary calibration method is more higher 3% than Primary calibration method. Finally, SOF which is included in soot particles is not totally removed so it have great influence on test result of counting efficiency and linearity.-
dc.language한국어-
dc.language.isoko-
dc.publisher한국액체미립화학회-
dc.relation.isPartOf한국액체미립화 학회지-
dc.subjectNano-particle()-
dc.subjectParticle Matters()-
dc.subjectCPC(Condensation Particle Counter)-
dc.subjectElectrometer-
dc.subjectCalibration-
dc.subjectValidation-
dc.subject나노입자-
dc.subject입자상물질-
dc.subject입자개수측정기-
dc.subject전위계-
dc.subject교정-
dc.subject검정-
dc.titleSpray type과 Soot type 입자발생기별 나노입자 개수농도분포 측정특성 비교-
dc.title.alternativeComparison on Nano-particle Number Measurement Characteristics for Different Particle Generators between Spray type and Soot type-
dc.typeArticle-
dc.type.rimsART-
dc.identifier.bibliographicCitation한국액체미립화 학회지, v.17, no.4, pp.185 - 191-
dc.identifier.kciidART001725238-
dc.description.journalClass2-
dc.citation.endPage191-
dc.citation.number4-
dc.citation.startPage185-
dc.citation.title한국액체미립화 학회지-
dc.citation.volume17-
dc.contributor.affiliatedAuthor이진욱-
dc.identifier.urlhttps://www.kci.go.kr/kciportal/ci/sereArticleSearch/ciSereArtiView.kci?sereArticleSearchBean.artiId=ART001725238-
dc.description.isOpenAccessN-
dc.subject.keywordAuthorNano-particle()-
dc.subject.keywordAuthorParticle Matters()-
dc.subject.keywordAuthorCPC(Condensation Particle Counter)-
dc.subject.keywordAuthorElectrometer-
dc.subject.keywordAuthorCalibration-
dc.subject.keywordAuthorValidation-
dc.subject.keywordAuthor나노입자-
dc.subject.keywordAuthor입자상물질-
dc.subject.keywordAuthor입자개수측정기-
dc.subject.keywordAuthor전위계-
dc.subject.keywordAuthor교정-
dc.subject.keywordAuthor검정-
dc.description.journalRegisteredClasskci-
Files in This Item
Go to Link
Appears in
Collections
College of Engineering > ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Jin wook photo

Lee, Jin wook
College of Engineering (School of Mechanical Engineering)
Read more

Altmetrics

Total Views & Downloads

BROWSE