Result of Proficiency Test and Comparison of Accuracy Using a European Spine Phantom among the Three Bone Densitometriesopen access

Authors
박애자최지혜강현박기정김하영김서화김덕윤박승환하용찬
Issue Date
2015
Publisher
대한골대사학회
Keywords
Bone density; Densitometry; Lumbar vertebrae; Reference standards
Citation
대한골대사학회지, v.22, no.2, pp 45 - 49
Pages
5
Journal Title
대한골대사학회지
Volume
22
Number
2
Start Page
45
End Page
49
URI
https://scholarworks.bwise.kr/cau/handle/2019.sw.cau/10931
DOI
10.11005/jbm.2015.22.2.45
ISSN
2287-6375
2287-7029
Abstract
Background: Although dual energy X-ray absorptiometry (DXA) is known to standard equipment for bone mineral density (BMD) measurements. Different results of BMD measurement using a number of different types of devices are difficult to use clinical practice. The purpose of this study was to evaluate discrepancy and standardizations of DXA devices from three manufactures using a European Spine Phantom (ESP). Methods: We calculated the accuracy and precision of 36 DXA devices from three manufacturers (10 Hologic, 16 Lunar, and 10 Osteosys) using a ESP (semi-anthropomorphic). The ESP was measured 5 times on each equipment without repositioning. Accuracy was assessed by comparing BMD (g/cm2 ) values measured on each device with the actual value of the phantom. Precision was assessed by the coefficient of variation (CVsd). Results: Lunar devices were, on average, 22%, 8.3%, and 5% overestimation for low (L1) BMD values, medium (L2) and high (L3) BMD values. Hologic devices were, on average, 6% overestimation for L1 BMD, and 5% and 6.2% underestimation for L2 and L3 BMD values. Osteosys devices was, on average, 12.7% (0.063 g/cm2 ), 6.3% (0.062 g/cm2 ), and 5% (0.075 g/ cm2 ) underestimation for L1, L2, and L3, respectively. The mean CVsd for L1-L3 BMD were 0.01%, 0.78%, and 2.46% for Lunar, Hologic, and Osteosys devices respectively. Conclusions: The BMD comparison in this study demonstrates that BMD result of three different devices are significant different between three devices. Differences of BMD between three devices are necessary to BMD standardization.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > College of Medicine > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Park, Seunghwan photo

Park, Seunghwan
의과대학 (의학부(임상-서울))
Read more

Altmetrics

Total Views & Downloads

BROWSE