Performance Characteristic of the beta-Human Chorionic Gonadotropin Assay on the Access 2 for Its Use in Doping Control
- Authors
- Kang, Min-Jung; Lee, Hyeim; Shin, Sang Mi; Yoo, Hye Hyun; Kwon, Oh-Seung; Jin, Changbae; Chung, Bong Chul
- Issue Date
- Oct-2010
- Publisher
- 대한화학회
- Keywords
- beta-hCG; Access 2; Doping control; Method evaluation; Confirmation method
- Citation
- Bulletin of the Korean Chemical Society, v.31, no.10, pp.2889 - 2892
- Indexed
- SCIE
SCOPUS
KCI
- Journal Title
- Bulletin of the Korean Chemical Society
- Volume
- 31
- Number
- 10
- Start Page
- 2889
- End Page
- 2892
- URI
- https://scholarworks.bwise.kr/erica/handle/2021.sw.erica/39456
- DOI
- 10.5012/bkcs.2010.31.10.2889
- ISSN
- 0253-2964
- Abstract
- The objective of this study was to validate performance characteristics of the Access 2 (Beckman coulter) system for hCG assays for use as a confirmation test for doping control. The Access 2 assay was linear up to 500 IU/L. The correlation coefficient was higher than 0.999, and the sensitivity of the linearity was 0.82. There were no false positive or false negative cases. LOD was 0.59 IU/L. The method was robust when performed by different people. Repeatability and reproducibility were below 7%. We compared reproducibility and recoveries of Access 2 and Elecsys 2010. Access 2 demonstrated higher reproducibility than Elecsys 2010. Recoveries (accuracy) of Access 2 were between 85 and 105%. Recoveries for Elecsys 2010 were between 91 and 104%.
- Files in This Item
-
Go to Link
- Appears in
Collections - COLLEGE OF PHARMACY > DEPARTMENT OF PHARMACY > 1. Journal Articles
![qrcode](https://api.qrserver.com/v1/create-qr-code/?size=55x55&data=https://scholarworks.bwise.kr/erica/handle/2021.sw.erica/39456)
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.