Evaluation of the planar inductive magnetic field sensors for metallic crack detections
- Authors
- Cha, Yu-Jung; Nam, Baekil; Kim, Jongryoul; Kim, Ki Hyeon
- Issue Date
- Jul-2010
- Publisher
- ELSEVIER SCIENCE SA
- Keywords
- Soft magnetic thin film; Thin film devices; Nondestructive testing; Magnetic field sensors; Inductive sensor
- Citation
- SENSORS AND ACTUATORS A-PHYSICAL, v.162, no.1, pp 13 - 19
- Pages
- 7
- Indexed
- SCI
SCIE
SCOPUS
- Journal Title
- SENSORS AND ACTUATORS A-PHYSICAL
- Volume
- 162
- Number
- 1
- Start Page
- 13
- End Page
- 19
- URI
- https://scholarworks.bwise.kr/erica/handle/2021.sw.erica/39673
- DOI
- 10.1016/j.sna.2010.06.003
- ISSN
- 0924-4247
- Abstract
- The planar inductive magnetic field sensors were fabricated by photolithography process. The sensor was composed of the planar coils (42 turns and 120 turns) with magnetic thin film as a core. To evaluate the sensor for crack detection, the nonmagnetic Al and magnetic FeC specimens were prepared with the circular and slit shaped artificial surface cracks. The alternative current (0.1 A to 1.4 A) with frequency range of 1.0-2.0 MHz was generated by the straight shaped driving Cu coil. In the case of 120 turns of planar inductive magnetic field sensor, the induced signals on crack positions exhibited the high intensities and stable signal to ratio as varying the driving current and frequency in comparison with those of 42 turns of planar inductive sensor. The measured output signals on FeC specimen with the micro-sized surface crack by the non-contacted scanning were converted to the image which was compared with that of the optical image, respectively. (C) 2010 Elsevier B.V. All rights reserved.
- Files in This Item
-
Go to Link
- Appears in
Collections - COLLEGE OF ENGINEERING SCIENCES > DEPARTMENT OF MATERIALS SCIENCE AND CHEMICAL ENGINEERING > 1. Journal Articles

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.