Radio frequency-compensated Langmuir probe with auxiliary double probes
- Authors
- Oh, Se-Jin; Oh, Seung-Ju; Chung, Chin-Wook
- Issue Date
- Sep-2010
- Publisher
- American Institute of Physics
- Keywords
- Langmuir probes; plasma sheaths
- Citation
- Review of Scientific Instruments, v.81, no.9, pp 1 - 4
- Pages
- 4
- Indexed
- SCI
SCIE
SCOPUS
- Journal Title
- Review of Scientific Instruments
- Volume
- 81
- Number
- 9
- Start Page
- 1
- End Page
- 4
- URI
- https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/174154
- DOI
- 10.1063/1.3478338
- ISSN
- 0034-6748
1089-7623
- Abstract
- A radio frequency (rf) compensation design using auxiliary double probes connected in parallel with a main measurement probe was developed for Langmuir probe diagnostics. This probe structure can reduce the sheath impedance of the main probe. In our probe design, the sheath capacitance of the probe can be increased and its sheath resistance can be decreased with increasing dc bias differential voltage between the auxiliary double probes. The I-V characteristic curve and electron energy distribution functions measured by our probe system had sufficient rf compensation performance in inductively coupled plasmas.
- Files in This Item
-
Go to Link
- Appears in
Collections - 서울 공과대학 > 서울 전기공학전공 > 1. Journal Articles

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.