Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Applicability of X-ray microscopy to non-contact tablet mass estimation

Authors
Kim, JiseokYim, ChewookKim, Yongkyun
Issue Date
Jun-2018
Publisher
SPRINGER
Keywords
Mass measurement; Tablet manufacturing; X-ray transmittance; Imaging plate detector; Volume measurement
Citation
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, v.316, no.3, pp.1033 - 1038
Indexed
SCIE
SCOPUS
Journal Title
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY
Volume
316
Number
3
Start Page
1033
End Page
1038
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/2359
DOI
10.1007/s10967-018-5867-4
ISSN
0236-5731
Abstract
The pharmaceutical tablet manufacturing process requires a tablet mass measuring step to ensure that all tablets have the prescribed dosage. The digital scales commonly used for that purpose require tens of seconds to stabilize, resulting in a substantial slow-down of the manufacturing process. The alternative visual approaches are fallible because of the potential existence of voids and cracks in the tablets. This paper uses an X-ray microscopy for mass measurement. Albeit with some constraints concerning tablet size and constituents, the approach can measure a tablet's mass in under one second, and may constitute, in the future, a viable and practical approach.
Files in This Item
Go to Link
Appears in
Collections
서울 공과대학 > 서울 원자력공학과 > 1. Journal Articles